  • How to Open and Close Housing (SOMNOcomfort and SOMNObalance Ventilator)
  • 9 steps

Flags

0

  • BackSOMNOcomfort and SOMNObalance

  • Full Screen

  • Options

  • History

  • Save to Favorites

  • Download PDF

  • Edit

  • Translate

  • Get Shareable Link

  • Embed This Guide

  • Notify Me of Changes

  • Stop Notifications

Introduction

What you need

Step 1

              How to Open and Close Housing (SOMNOcomfort and SOMNObalance Ventilator)               
  • Remove the covering cap. (pg. 23)

Remove the covering cap. (pg. 23)

1024

Step 2

  • Place the unit upside down on a non-slip surface and remove the six screws. (pg. 23)

Place the unit upside down on a non-slip surface and remove the six screws. (pg. 23)

Step 3

  • Remove the lower part of the housing with a swiveling motion away to the side. Note the SOMNOcomfort series uses a shorter circuit board than the balance series ventilator. (pg. 23)

Remove the lower part of the housing with a swiveling motion away to the side. Note the SOMNOcomfort series uses a shorter circuit board than the balance series ventilator. (pg. 23)

Step 4

  • Remove the connector (X1) from the circuit board.
  • Pull the connecting cable (64) from its mount.
  • Now set aside the lower part of the housing. (pg. 23)

Remove the connector (X1) from the circuit board.

Pull the connecting cable (64) from its mount.

Now set aside the lower part of the housing. (pg. 23)

Step 5

  • Next, how to close the compartment and reverse the process.
  • Hold the lower part of the housing to the upper part at an angle
  • Attach the connector (X1) to the main circuit board. When doing so the connector’s nub has to point towards the lock. (pg. 24)

Next, how to close the compartment and reverse the process.

Hold the lower part of the housing to the upper part at an angle

Attach the connector (X1) to the main circuit board. When doing so the connector’s nub has to point towards the lock. (pg. 24)

Step 6

  • Properly align the main circuit board inside the unit.
  • Attach the connecting cable (64) to its mount (pg. 24)

Properly align the main circuit board inside the unit.

Attach the connecting cable (64) to its mount (pg. 24)

Step 7

  • Place the lower part of the housing on the upper part (47). Please make sure the main circuit board (44) is aligned properly, and that the connecting cable (64) is not being squeezed anywhere. (pg. 24)

Place the lower part of the housing on the upper part (47). Please make sure the main circuit board (44) is aligned properly, and that the connecting cable (64) is not being squeezed anywhere. (pg. 24)

Step 8

  • Now reattach the upper part of the housing by tightening the six screws. (pg. 24)

Now reattach the upper part of the housing by tightening the six screws. (pg. 24)

Step 9

  • Insert the covering cap. (pg. 24)

Insert the covering cap. (pg. 24)

Now, you can properly assemble and reassemble the SOMNOcomfort and SOMNObalance ventilators in order to perform and necessary repairs.

Cancel: I did not complete this guide.

Author

                                      with 2 other contributors 

                    Diya Jamwal                     

Member since: 11/07/2018

553 Reputation

                                      3 Guides authored                  



                       Badges:
                       5







                                                        +2 more badges                           

Team

                       Cal Poly, Team S1-G3, Zimmerman Spring 2020                        

                                                  Member of Cal Poly, Team S1-G3, Zimmerman Spring 2020 



                    CPSU-ZIMMERMAN-S20S1G3                     


                                            5 Members                     


                                            10 Guides authored