Difficulty
Moderate
Steps
8
Time Required
20 minutes
Sections
1
- Electrochemical Sensors
- 8 steps
Flags
0
BackMedical Device
Full Screen
Options
History
Save to Favorites
Download PDF
Edit
Translate
Get Shareable Link
Embed This Guide
Notify Me of Changes
Stop Notifications
Introduction
What you need
Step 1
Electrochemical Sensors
- Confirm which sensor you are intending to replace. Make note of it.
- Nitrile gloves should be worn for the entire procedure.
Confirm which sensor you are intending to replace. Make note of it.
Nitrile gloves should be worn for the entire procedure.
1024
Step 2
- Place the monitor face down on a clean surface.
- Use the Phillips #0 screwdriver to remove the single 3 mm Phillips screw on the belt clip.
Place the monitor face down on a clean surface.
Use the Phillips #0 screwdriver to remove the single 3 mm Phillips screw on the belt clip.
Step 3
- Starting at the bottom of the case, pull the outer case away from the main body of the monitor.
- Continue pulling out and away until the entire outer case has been removed.
Starting at the bottom of the case, pull the outer case away from the main body of the monitor.
Continue pulling out and away until the entire outer case has been removed.
Step 4
- With the Phillips #0 screwdriver, remove all four 3 mm screws from the rear pump/sensor cover.
With the Phillips #0 screwdriver, remove all four 3 mm screws from the rear pump/sensor cover.
Step 5
- Remove the outer pump/sensor housing cover and set it aside.
Remove the outer pump/sensor housing cover and set it aside.
Step 6
- Carefully grip and twist the outer edge of the sensor to remove the sensor from its insertion point.
- Be careful not to twist too aggressively as this could damage the sensor attachment point in the monitor.
Carefully grip and twist the outer edge of the sensor to remove the sensor from its insertion point.
Be careful not to twist too aggressively as this could damage the sensor attachment point in the monitor.
Step 7
- Grip the replacement sensor on the outer edge.
- Insert the new sensor in a straight downward orientation.
- Be careful to line up the prongs with their insertion/attachment point in the monitor.
Grip the replacement sensor on the outer edge.
Insert the new sensor in a straight downward orientation.
Be careful to line up the prongs with their insertion/attachment point in the monitor.
Step 8
- Reassemble the unit in reverse order.
- After reassembly, power the unit on to ensure that the monitor recognizes the new sensor.
- Calibrate the sensor per the manufacturer’s instructions.
Reassemble the unit in reverse order.
After reassembly, power the unit on to ensure that the monitor recognizes the new sensor.
Calibrate the sensor per the manufacturer’s instructions.
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
Author
with 2 other contributors
Steven E Boughey
Member since: 01/04/2021
197 Reputation
1 Guide authored
Badges:
3
Team
Oregon Institute of Technology, Team S2-G2, Rauch Winter 2021
Member of Oregon Institute of Technology, Team S2-G2, Rauch Winter 2021
OIT-RAUCH-W21S2G2
1 Member
1 Guide authored