Introduction

What you need

Step 1

              Electrochemical Sensors               
  • Confirm which sensor you are intending to replace. Make note of it.
  • Nitrile gloves should be worn for the entire procedure.

Confirm which sensor you are intending to replace. Make note of it.

Nitrile gloves should be worn for the entire procedure.

1024

Step 2

  • Place the monitor face down on a clean surface.
  • Use the Phillips #0 screwdriver to remove the single 3 mm Phillips screw on the belt clip.

Place the monitor face down on a clean surface.

Use the Phillips #0 screwdriver to remove the single 3 mm Phillips screw on the belt clip.

Step 3

  • Starting at the bottom of the case, pull the outer case away from the main body of the monitor.
  • Continue pulling out and away until the entire outer case has been removed.

Starting at the bottom of the case, pull the outer case away from the main body of the monitor.

Continue pulling out and away until the entire outer case has been removed.

Step 4

  • With the Phillips #0 screwdriver, remove all four 3 mm screws from the rear pump/sensor cover.

With the Phillips #0 screwdriver, remove all four 3 mm screws from the rear pump/sensor cover.

Step 5

  • Remove the outer pump/sensor housing cover and set it aside.

Remove the outer pump/sensor housing cover and set it aside.

Step 6

  • Carefully grip and twist the outer edge of the sensor to remove the sensor from its insertion point.
  • Be careful not to twist too aggressively as this could damage the sensor attachment point in the monitor.

Carefully grip and twist the outer edge of the sensor to remove the sensor from its insertion point.

Be careful not to twist too aggressively as this could damage the sensor attachment point in the monitor.

Step 7

  • Grip the replacement sensor on the outer edge.
  • Insert the new sensor in a straight downward orientation.
  • Be careful to line up the prongs with their insertion/attachment point in the monitor.

Grip the replacement sensor on the outer edge.

Insert the new sensor in a straight downward orientation.

Be careful to line up the prongs with their insertion/attachment point in the monitor.

Step 8

  • Reassemble the unit in reverse order.
  • After reassembly, power the unit on to ensure that the monitor recognizes the new sensor.
  • Calibrate the sensor per the manufacturer’s instructions.

Reassemble the unit in reverse order.

After reassembly, power the unit on to ensure that the monitor recognizes the new sensor.

Calibrate the sensor per the manufacturer’s instructions.

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

Author

                                      with 2 other contributors 

                    Steven E Boughey                     

Member since: 01/04/2021

197 Reputation

                                      1 Guide authored                  



                       Badges:
                       3

Team

                       Oregon Institute of Technology, Team S2-G2, Rauch Winter 2021                        

                                                  Member of Oregon Institute of Technology, Team S2-G2, Rauch Winter 2021 



                    OIT-RAUCH-W21S2G2                     


                                            1 Member                     


                                            1 Guide authored