NOTE:
- Replace the oxygen cell with genuine Hamilton Medical parts only; otherwise, oxygen measurement will not function. To prevent a permanent alarm make sure to use special Hamilton Medical oxygen cells only.To prevent leakage within the ventilator, make sure an oxygen cell is installed at all times, even if you use an external monitor or disable oxygen monitoring.
What you need
Step 1
Oxygen Cells
- Flip device around until the back panel is exposed.
Flip device around until the back panel is exposed.
1024
Step 2
- Remove the screw in the bottom center of the panel, and then take the panel off.
Remove the screw in the bottom center of the panel, and then take the panel off.
Step 3
- Disconnect the oxygen cell by removing the tube.
- Then twist the cell counter-clockwise and pull to remove the oxygen cell.
- Install the new oxygen cell and put the back cover on.
Disconnect the oxygen cell by removing the tube.
Then twist the cell counter-clockwise and pull to remove the oxygen cell.
Install the new oxygen cell and put the back cover on.
Step 4
- After the cell is installed. Run the oxygen cell calibration.
- Oxygen calibration guide can be found in 3-11 of the Hamilton-T1 operations manual.
After the cell is installed. Run the oxygen cell calibration.
Oxygen calibration guide can be found in 3-11 of the Hamilton-T1 operations manual.
After the installation and calibration of the oxygen cell. The Hamilton-T1 Ventilator is ready to operate.
Cancel: I did not complete this guide.
Attached Documents
- HAMILTON-T1 Operator’s Manual SW2.1x EN
- PDF - 4.17 mb
- View
HAMILTON-T1 Operator’s Manual SW2.1x EN
PDF - 4.17 mb
Author
with 1 other contributor
Tyler
Member since: 04/09/2020
114 Reputation
1 Guide authored
Badges:
4
+1 more badge
Team
Cal Poly, Team S11-G24, Paton Spring 2020
Member of Cal Poly, Team S11-G24, Paton Spring 2020
CPSU-PATON-S20S11G24
3 Members
3 Guides authored