Introduction
What you need
Step 1
Rail
- Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.
Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.
1024
Step 2
- Remove the Rail (B).
Remove the Rail (B).
Step 3
- The Standoff Brackets (C) remain with the Main Cover.
The Standoff Brackets (C) remain with the Main Cover.
Step 4
- Place aside in a safe location.
Place aside in a safe location.
Step 5
Main Cover
- Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.
Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.
Step 6
Step 7
Safety Valve Block
- Remove Safety Valve Block
- Disconnect the cable from the Ambient Valve Solenoid
- Disconnect the Shaped Hose (A)
- Remove the 2 Hex Allen Screws (B)
- Remove the Safety Valve Block
- To remove Ambient Valve Solenoid: Unscrew the Ambient Valve Solenoid (C) form the Safety Valve Block
Remove Safety Valve Block
Disconnect the cable from the Ambient Valve Solenoid
Disconnect the Shaped Hose (A)
Remove the 2 Hex Allen Screws (B)
Remove the Safety Valve Block
To remove Ambient Valve Solenoid: Unscrew the Ambient Valve Solenoid (C) form the Safety Valve Block
To reassemble your device, follow these instructions in reverse order.