Difficulty

Moderate

Steps

8

Time Required

                          15 minutes            

Sections

4

  • Rail
  • 4 steps
  • Main Cover
  • 2 steps
  • Safety Valve Block
  • 1 step
  • Inspiratory Valve
  • 1 step

Flags

0

  • BackHamilton G5

  • Full Screen

  • Options

  • History

  • Save to Favorites

  • Download PDF

  • Edit

  • Translate

  • Get Shareable Link

  • Embed This Guide

  • Notify Me of Changes

  • Stop Notifications

Introduction

What you need

Step 1

              Rail               
  • Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.

Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.

1024

Step 2

  • Remove the Rail (B).

Remove the Rail (B).

Step 3

  • The Standoff Brackets (C) remain with the Main Cover.

The Standoff Brackets (C) remain with the Main Cover.

Step 4

  • Place aside in a safe location.

Place aside in a safe location.

Step 5

              Main Cover               
  • Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.

Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.

Step 6

Step 7

              Safety Valve Block               
  • Remove Safety Valve Block
    1. Disconnect the cable from the Ambient Valve Solenoid
    1. Disconnect the Shaped Hose (A)
    1. Remove the 2 Hex Allen Screws (B)
    1. Remove the Safety Valve Block
  • To remove Ambient Valve Solenoid: Unscrew the Ambient Valve Solenoid (C) form the Safety Valve Block

Remove Safety Valve Block

  1. Disconnect the cable from the Ambient Valve Solenoid

  2. Disconnect the Shaped Hose (A)

  3. Remove the 2 Hex Allen Screws (B)

  4. Remove the Safety Valve Block

To remove Ambient Valve Solenoid: Unscrew the Ambient Valve Solenoid (C) form the Safety Valve Block

Step 8

              Inspiratory Valve               
  • Removing the Inspiration Valve
    1. Disconnect the tubing and cables form the Inspiration Valve
    1. Disconnect the Shaped Hose (A)
    1. Remove 2 Hex Allen screws (B) M5x60
    1. Move the Inspiration Valve forward and then up
    1. Remove the Inspiration Valve

Removing the Inspiration Valve

  1. Disconnect the tubing and cables form the Inspiration Valve

  2. Remove 2 Hex Allen screws (B) M5x60

  3. Move the Inspiration Valve forward and then up

  4. Remove the Inspiration Valve

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

Attached Documents

  • Hamilton G5 Ventilator Service Manual.pdf - iFixit.pdf
  • PDF - 129.98 kb
  • View

Hamilton G5 Ventilator Service Manual.pdf - iFixit.pdf

PDF - 129.98 kb

Author

                                      with 2 other contributors 

                    William                     

Member since: 04/09/2020

145 Reputation

                                      2 Guides authored                  



                       Badges:
                       2

Team

                       Cal Poly, Team S7-G25, Paton Spring 2020                        

                                                  Member of Cal Poly, Team S7-G25, Paton Spring 2020 



                    CPSU-PATON-S20S7G25                     


                                            6 Members                     


                                            11 Guides authored                     

William - May 31, 2020

Reply

wow great manual! very helpful and informative!