Introduction
What you need
Step 1
Rail
- Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.
Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.
1024
Step 2
- Remove the Rail (B).
Remove the Rail (B).
Step 3
- The Standoff Brackets (C) remain with the Main Cover.
The Standoff Brackets (C) remain with the Main Cover.
Step 4
- Place aside in a safe location.
Place aside in a safe location.
Step 5
Main Cover
- Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.
Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.
Step 6
Step 7
Expiratory Valve
- Disconnect 2 cables from the magnet of the Expiratory Valve.
Disconnect 2 cables from the magnet of the Expiratory Valve.
Step 8
- Remove the Expiratory Valve Cover (A) and Silicon Membrane (A) (if not already removed).
Remove the Expiratory Valve Cover (A) and Silicon Membrane (A) (if not already removed).
Step 9
- Remove 3 Hex Allen screws (B) M3x8
Remove 3 Hex Allen screws (B) M3x8
Step 10
- Remove the Expiratory Valve.
Remove the Expiratory Valve.
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
Author
with 1 other contributor
Sierra Saenz
Member since: 04/09/2020
366 Reputation
3 Guides authored
Badges:
6
+3 more badges
Team
Cal Poly, Team S7-G25, Paton Spring 2020
Member of Cal Poly, Team S7-G25, Paton Spring 2020
CPSU-PATON-S20S7G25
6 Members
11 Guides authored