Introduction
PLEASE NOTE: This repair guide was developed by the iFixit team based on Hamilton’s own service manual. Neither iFixit nor this repair guide is endorsed by or affiliated with Hamilton.
Replacing the air inlet filter is discussed in:
- chapter 9.3.1 “Maintenance - servicing the air intake and fan filters” of user manualchapter 11.4.2 “Components Removal/Assembly - HEPA Filter” of service manual
Step 1
shut down the ventilator
- Press and quickly release the power switch to open standby window in the screen.
- Activate the standby mode by pressing the Active standby button in the screen.
- Press and hold the switch for greater than 3 seconds to shut down the ventilator.
- If there is a technical fault, press and hold the switch for greater than 10 seconds.
Press and quickly release the power switch to open standby window in the screen.
Activate the standby mode by pressing the Active standby button in the screen.
Press and hold the switch for greater than 3 seconds to shut down the ventilator.
If there is a technical fault, press and hold the switch for greater than 10 seconds.
1024
Step 2
disconnect from power
- Disconnect the ventilator from the AC or DC power.
Disconnect the ventilator from the AC or DC power.
Step 3
remove filter cover panel
- The filter cover panel is attached with 2 latches.
- Detach top latches, swivel away from the rear and remove the cover panel.
The filter cover panel is attached with 2 latches.
Detach top latches, swivel away from the rear and remove the cover panel.
Step 4
replace filters
- Remove dust filters.
- Raise the HEPA Filter Latch.
- Use the handle to pull the HEPA Filter from the holder and replace as required.
- Install new dust filters or wash the existing filter in a mild soap solution, rinse, dry and reinstall.
Remove dust filters.
Raise the HEPA Filter Latch.
Use the handle to pull the HEPA Filter from the holder and replace as required.
Install new dust filters or wash the existing filter in a mild soap solution, rinse, dry and reinstall.
- Place back the cover panelConnect the ventilator back to the AC or DC power.Power on the ventilator with the power switch.Run the preoperational check.